Verify before you connect
Every assembly is checked the same way: power, gas source, circuit integrity, then a functional test against a test lung.
- Ventilator pre-use check: leak test at 30 cm H2O, verify delivered volume against set volume on a test lung, confirm oxygen analyzer accuracy at 21% and 100%, confirm alarms annunciate.
- Humidifier: water level between the min and max lines, temperature probe at the patient wye, heated-wire circuit set 2–3 degrees above chamber temperature to limit rainout.
- Manual resuscitator: check valve function, mask seal, oxygen reservoir attached, and flow of 10–15 L/min before it is ever needed.
- Suction: verify wall vacuum at −80 to −120 mm Hg for adults with the tubing occluded, not open.
A device that has not been functionally tested against a test lung has not been checked — a visual inspection catches almost nothing.